- Designed as per ICH guidelines to meet, WHO and USFDA requirement
- Fulfills storage conditions of 25⁰C – 60% Rh, 40⁰C – 75% Rh, 30⁰C – 65% Rh, 25⁰C – 40% Rh
- Precise monitoring of temperature and humidity conditions
- Comply with GMP requirements
|
- Powerful fan motor for forced air circulation to maintain uniform conditions inside chamber within permissible limit
- Heating by long life SS tubular heaters
- Validation protocol with IQ, OQ, DQ Documentation as per ICH guidelines
|
| |
 |
- Microprocessor controller with LCD display & capacitance type sensor for precise control of temperature and humidity
4 Inch attractive LCD display for direct display of humidity & temperature
- Battery Backup for Controller - 12 V, 7.2 Ah sealed maintenance free, Fully charged battery runs for more than eight hours
- Automatic Calibration of set parameters
|
- Self-Diagnosis for errors
- Power failure and resumption recorded with date and time
- Machine filled PUF insulation to eliminate void pockets
- Unique design of thermal barrier for better energy efficiency
- Hermetically sealed Compressor with CFC free refrigerant
|
| |
 |
- Histogram format of 24 hours temperature and humidity recording
- Settable High / low alarm points with date and time
- Memory for storage of up to 1000 records
- Centronic interface to record Temperature, Humidity, Date, & Time by attaching Dot Matrix printer with adjustable print interval
- Chamber calibration port on side
|
- Adjustable trays
- PC communication with RS-485 and software complying to 21 CFR part 11 as per USFDA guidelines (Optional)
- Data logger for 8 Point (4 + 4) temperature / humidity recording with sensor, printer interface and software (Optional)
|
| |
 |
- Low water level alarm and power cutoff of boiler tank heater
- High temperature safety cut off & alarms for high/low set parameters
- Electrical circuit breaker
|
- Time delay for compressor switch on
- Overload cut off relay for compressor
|
| |
| Features of Software complying to 21 CFR part 11 |
- Multiple level alpha numerical password with password ageing
- Electronic signature and menu entered comment
- Secured audit trail report
- Alarm report with mention of alarm condition
- Mean Kinetic Temperature able to be calculated for any days & alarm deviation report
|
- Reports in graphical and tabular form
- Data Stored in PC server with Lan connectivity and can be monitored on any PC with password authentication
- Multi chamber data acquisition on single software
|
| |
 |
| |
| Technical data |
| |
SC-6 Plus |
SC-10 Plus |
SC-12 Plus |
SC-16 Plus |
SC-19 Plus |
| Internal Volume (Liters) |
200 |
280 |
370 |
450 |
540 |
| Number of Shelves |
2 |
3 |
4 |
5 |
6 |
| Internal Dimensions W x D x H (mm) |
660 x 655 x 450 |
660 x 655 x 650 |
660 x 655 x 850 |
660 x 655 x 1050 |
660 x 655 x 1250 |
| External Dimensions W x D x H (mm) |
800 x 1095 x 1020 |
800 x 1095 x 1220 |
800 x 1095 x 1420 |
800 x 1095 x 1620 |
800 x 1095 x 1820 |
| Temperature Range & Accuracy |
10°C to 60°C, ± 1°C |
| Humidity Range (Rh) & Accuracy |
40% to 95%, ± 3% |
| Temperature & Humidity control |
Microprocessor with Capacitance type sensor |
| Display |
4" LCD Screen, Large size Display for ease of reading |
| Power Failure Alarm |
Audio Visual Alarm |
| Door Open Alarm |
Audio Alarm in case door open for over one minute |
| Humidity/Temperature Variation Alarm |
Set Humidity ± 3% Rh, Set Temperature ± 2°C, Audio Visual Alarm |
| Illumination |
8 Watts Fluorescent Lamp |
| Internal Body Material |
Stainless Steel - 316 grade |
| External Body Material |
Stainless Steel - 304 grade |
| Insulation |
70 mm minimum for Body & 80 mm for Door, CFC free polyurethane foam |
| Noise Level |
Less Than 65 db(A) |
| Recommended Voltage Stabilizer |
VS-04 |
| Electrical |
220-240 volts, 50Hz, Single Phase |
|